The source is a 2 mm thick plexiglass plate. The ceramic active part with the embedded radionuclide preparation is placed in the center of the body, the sealing is made with epoxy resin. The diameter of the active part is 2 mm, the height is 0.5 mm.
Radionuclide | Half-life, days | Photon energy, keV | Photons per 100 disintegration | Activity, kBq * | Impurities, % ** | Coefficients of the photon radiation transmission;*** | ||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
22Na | 950.57 ± 0.23 | 1274.54 | 0.99940 ± 0.00014 | 5 – 1000 | ≤ 0.01 | 0.988 | ||||||||||
54Mn | 312.29 ± 0.26 | 834.848 | 0.999746 ± 0.000011 | 5 – 1000 | ≤ 0.15 | 0.985 | ||||||||||
57Co | 271.80 ± 0.05 | 122.061 | 0.8551 ± 0.0006 | 5 – 3700 | ≤ 0.15 | 0.967 | ||||||||||
60Co | 1925.23 ± 0.27 | 1332.5 | 0.999826 ± 0.000006 | 5 – 3700 | ≤ 0.1 | 0.988 | ||||||||||
88Y | 106.625 ± 0.024 | 1836.066 | 0.9938 ± 0.0003 | 5 – 1000 | ≤ 0.1 | 0.991 | ||||||||||
109Cd | 461.4 ± 1.2 | 88.034 | 0.03626 ± 0.00020 | 5 – 1000 | ≤ 0.001 | 0.965 | ||||||||||
133Ba | 3848.7 ± 1.2 | 356.014 | 0.6205 ± 0.0019 | 5 – 3700 | ≤ 0.001 | 0.977 | ||||||||||
137Cs | 10990 ± 40 | 661.66 | 0.8499 ± 0.0020 | 5 – 1000 | ≤ 0.05 | 0.983 | ||||||||||
152Eu | 4941 ± 7 | 1408.006 | 0.2085 ± 0.0009 | 5 – 1000 | ≤ 0.15 | 0.988 | ||||||||||
241Am | 157850 ± 230 | 59.537 | 0.3578 ± 0.0009 | 5 – 370 | ≤ 0.05 | 0.962 | ||||||||||
* The nominal value of the radionuclide activity in the source within the specified limits is determined by the consumer when the source is ordered. Tolerance on activity -10 ÷ +30 %. ** The ratio of the total activity of impurity radionuclides to the activity of the basic radionuclide in the source. *** Coefficients of the photon radiation transmission in source:
Recommended working life
|
OSGI-P sources are used for efficiency and energy calibration of gamma-spectrometric equipment, testing of ionizing chambers and scintillation counters. Robust design is suitable for “in field” calibrations and tests.
Source is a transparent disc with diameter 25 mm and height 3 mm. Radionuclide is deposited on active part made of light ceramic pellet. Active part is mounted in center of acrylic capsule and sealed with epoxy resin. Dimensions of active part are 2 mm in diameter and 1 mm in height.
Radionuclide | Half-life, days | Photon energy, keV | Photons per 100 disintegration | Activity, kBq * | Impurities, % ** | Coefficients of the photon radiation transmission *** | ||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
22Na | 950.57 ± 0.23 | 1274.54 | 0.99940 ± 0.00014 | 5 – 1000 | ≤ 0.01 | 0.988 | ||||||||||
44Ti | 21915 ± 400 | 67.884 | 0.930 ± 0.020 | 5 – 100 | ≤ 0.001 | 0.963 | ||||||||||
54Mn | 312.29 ± 0.26 | 834.848 | 0.999746 ± 0.000011 | 5 – 1000 | ≤ 0.15 | 0.985 | ||||||||||
57Co | 271.80 ± 0.05 | 122.061 | 0.8551 ± 0.0006 | 5 – 3700 | ≤ 0.15 | 0.967 | ||||||||||
60Co | 1925.23 ± 0.27 | 1332.5 | 0.999826 ± 0.000006 | 5 – 3700 | ≤ 0.1 | 0.988 | ||||||||||
65Zn | 243.86 ± 0.20 | 1115.55 | 0.5060 ± 0.0022 | 5 – 1000 | ≤ 0.2 | 0.987 | ||||||||||
88Y | 106.625 ± 0.024 | 1836.066 | 0.9938 ± 0.0003 | 5 – 1000 | ≤ 0.1 | 0.991 | ||||||||||
109Cd | 461.4 ± 1.2 | 88.034 | 0.03626 ± 0.00020 | 5 – 1000 | ≤ 0.001 | 0.965 | ||||||||||
113Sn | 115.09 ± 0.04 | 391.71 | 0.6494 ± 0.0017 | 5 – 100 | ≤ 0.01 | 0.978 | ||||||||||
133Ba | 3848.7 ± 1.2 | 356.014 | 0.6205 ± 0.0019 | 5 – 3700 | ≤ 0.001 | 0.977 | ||||||||||
134Cs | 753.5 ± 1.0 | 604.7 | 0.97650 ± 0.00018 | 5 – 300 | ≤ 0.01 | 0.982 | ||||||||||
137Cs | 10990 ± 40 | 661.66 | 0.8499 ± 0.0020 | 5 – 1000 | ≤ 0.05 | 0.983 | ||||||||||
139Ce | 137.642 ± 0.020 | 165,86 | 0.7990 ± 0.0004 | 5 – 1000 | ≤ 0.1 | 0.969 | ||||||||||
152Eu | 4941 ± 7 | 1408.006 | 0.2085 ± 0.0009 | 5 – 1000 | ≤ 0.15 | 0.988 | ||||||||||
153Gd | 240.4 ± 1 | 97.431 | 0.29 | 5 – 500 | ≤ 0.2 | 0.965 | ||||||||||
207Bi | 11800 ± 300 | 569.7 | 0.9776 ± 0.0003 | 5 – 100 | ≤ 0.2 | 0.982 | ||||||||||
228Th | 698.6 ± 0.23 | 2614 | 0.99754 ± 4 | 5 – 50 | ≤ 0.2 | 0.998 | ||||||||||
241Am | 157850 ± 230 | 59.537 | 0.3578 ± 0.0009 | 5 – 370 | ≤ 0.05 | 0.962 | ||||||||||
243Am | 2692000 ± 8000 | 74.66 | 0.672 | 5 – 50 | < 5 | 0.964 | ||||||||||
* The nominal value of the radionuclide activity in the source within the specified limits is determined by the consumer when the source is ordered. Tolerance on activity -10 ÷ +30 %. ** The ratio of the total activity of impurity radionuclides to the activity of the basic radionuclide in the source. *** Coefficients of the photon radiation transmission in OSGI-P source:
Recommended working life
|
Gamma standard sources are used as working standards for verification and graduation of photon measuring equipment, for certification of measuring methods and measurement accuracy control of photon emission. In particular, the set is intended for testing and calibration of precision instruments, spectrometers and radiometers, ionization chambers and scintillation counters.
The source is produced in the form of a flat round capsule made of aluminum, 25 mm in diameter (29 mm by special order) and 3 mm thick. The source active part of less than 3 mm in diameter is embedded between two polyimide foils with total thickness of 100 ± 10 μm.
Radionuclide | Decay mode | Half-life, days | Photon energy, keV | Photons per 100 disintegration | Activity, kBq * | Equivalent dose rate, μSv/h ** | |||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
22Na | EC | 950.57 ± 0.23 | 511 (annihil) 1274.537 ± 0.003 |
1.798 ± 0.002 0.99940 ± 0.00014 |
1 – 1000 | ≤ 35.0 | |||||||||||||||||||||
44Ti | EC | 21900 ± 15 |
68.868 ± 0.014 |
0.930 ± 0.0015 0.964 ± 0.0011 184.54 ± 0.10 0.999 ± 0.004 |
1 – 300 | ≤ 10.2 | |||||||||||||||||||||
54Mn | EC | 312.29 ± 0.26 | 834.838 ± 0.005 | 0.999746 ± 0.000011 | 1 – 1000 | ≤ 14.0 | |||||||||||||||||||||
55Fe | EC | 1002.7 ± 2.3 | (Mn Kα) 5,8988 (Mn Kβ) 6.49-6.54 |
0.1656 ± 0.0027 0.0340 ± 0.007 |
1 – 1000 | ≤ 0.02 | |||||||||||||||||||||
57Co | EC | 271.80 ± 0.05 | 14.41295 ± 0.00031 122.06065 ± 0.00012 136.47356 ± 0.00029 |
0.0915 ± 0.0017 0.8551 ± 0.0006 0.1071 ± 0.0015 |
1 – 1000 | ≤ 1.7 | |||||||||||||||||||||
60Co | β- | 1925.23 ± 0.27 | 1173.228 ± 0.003 1332.492 ± 0.004 |
0.9985 ± 0.0003 0.999826 ± 0.000006 |
1 – 500 | ≤ 19.0 | |||||||||||||||||||||
65Zn | EC | 243.86 ± 0.20 | 511 (annihil) 1115.539 ± 0.002 |
0.0284 ± 0.0004 0.5060 ± 0.0022 |
1 – 1000 | ≤ 10.0 | |||||||||||||||||||||
88Y | EC | 106.625 ± 0.024 | 898.036 ± 0.004 1836.052 ± 0.013 |
0.9390 ± 0.0023 0.9938 ± 0.0003 |
1 – 1000 | ≤ 38.8 | |||||||||||||||||||||
109Cd | EC | 461.4 ± 1.2 | 88.0336 ± 0.0011 | 0.03626 ± 0.00020 | 1 – 1000 | ≤ 0.04 | |||||||||||||||||||||
113Sn | EC | 115.09 ± 0.04 | 391.698 ± 0.003 | 0.6494 ± 0.0017 | 1 – 500 | ≤ 2.2 | |||||||||||||||||||||
133Ba | EC | 3848.7 ± 1.2 | 80.9979 ± 0.0011 276.3989 ± 0.0012 302.8508 ± 0.0005 356.0129 ± 0.0007 383.8485 ± 0.0012 |
0.329 ± 0.003 0.0716 ± 0.0005 0.1834 ± 0.0013 0.6205 ± 0.0019 0.0894 ± 0.0006 |
1 – 1000 | ≤ 8.8 | |||||||||||||||||||||
134Cs | β- | 753.5 ± 1.0 | 569.327 ± 0.003 604.720 ± 0.003 795.83 ± 0.03 |
0.1538 ± 0.0004 0.97650 ± 0.00018 0.855 ± 0.003 |
1 – 300 | ≤ 7.5 | |||||||||||||||||||||
137Cs | β- | 10990 ± 40 | 661.657 ± 0.003 | 0.8499 ± 0.0020 | 1 – 1000 | ≤ 10.0 | |||||||||||||||||||||
139Ce | EC | 137.642 ± 0.020 | 165.8575 ± 0.0011 | 0.7990 ± 0.0004 | 1 – 1000 | ≤ 6.3 | |||||||||||||||||||||
152Eu | EC | 4941 ± 7 | 121.7817 ± 0.0003 344.2785 ± 0.0012 778.9045 ± 0.0024 964.072 ± 0.018 1085.837 ± 0.010 1112.076 ± 0.003 1408.013 ± 0.003 |
0.2841 ± 0.0013 0.2658 ± 0.0012 0.1296 ± 0.0006 0.1462 ± 0.0006 0.1013 ± 0.0006 0.1340 ± 0.0006 0.2085 ± 0.0009 |
1 – 1000 | ≤ 20.0 | |||||||||||||||||||||
153Gd | EC | 240.4 ± 1 | 97.431 ± 0.021 103.18012 ± 0.00017 |
0.29 0.211 ± 0.006 |
1 – 500 | ≤ 1.0 | |||||||||||||||||||||
207Bi | EC | 11800 ± 300 | 569.698 ± 0.002 1063.656 ± 0.003 1770.228 ± 0.009 |
0.9776 ± 0.0003 0.7458 ± 0.0049 0.0687 ± 0.0002 |
1 – 100 | ≤ 2.6 | |||||||||||||||||||||
228Th | α | 670 ± 2 | 39.86 ÷ 2614.53 | 1.03 ÷ 36.06 | 1 – 100 | ≤ 0.3 | |||||||||||||||||||||
241Am | α | 157850 ± 230 | 26.3446 ± 0.0002 59.5409 ± 0.0001 |
0.0240 ± 0.0003 0.3578 ± 0.0009 |
1 – 300 | ≤ 0.12 | |||||||||||||||||||||
243Am | α | 2692000 ± 8000 | 43.53 ± 0.02 74.66 ± 0.02 |
0,0590 ± 0.0015 0.672 |
1 – 50 | ≤ 0.04 | |||||||||||||||||||||
* Other activities are available on request. The nominal value of the radionuclide activity in the source within the specified limits is determined by the consumer when the source is ordered. Tolerance on activity ±20 % Equivalent dose rate - equivalent dose rate of photon radiation at a distance of 0.1 m. Limit of error of measuring result for confidence probability р = 0.95: 7, 3 or 1.5 % — at certification, attestation in accredited metrological centers or attestation as working standards of the 0-th category, respectively. ISO classification: С35242 ** Estimated for maximum activity at 10 cm from source surface. Recommended working life and Intercalibration interval *
|
The source is produced in the form of a flat round capsule made of aluminum, 25 mm in diameter (29 mm by special order) and 3 mm thick. The source active part of less than 3 mm in diameter is embedded between two polyimide foils with total thickness of 100 ± 10 μm. The active spot of the source contains 4 nuclides:
88Y – 370 kBq, 133Ba – 37 kBq, 152Eu – 37 kBq, 241Am – 37 kBq.
At least 20 energy lines are required for efficiency calibration of semiconductor detector of HPGe and Ge(Li) types. Source is designed for 2 years useful work.
All calibration data are acquired in one run. This reduces systematic mistakes, while all components are measured at the same time and exposition.
Time saving compare to calibration with single nuclide sources.
используйте горизонтальную прокрутку
Radionuclide | Decay mode | Half-life, days | Photon energy, keV | Photons per 100 disintegration | Activity, kBq * | Exposure dose rate, μSv/h ** |
---|---|---|---|---|---|---|
88Y + | EC | 106.625 ± 0.024 | 898.036 ± 0.004 1836.052 ± 0.013 |
0.9390 ± 0.0023 0.9938 ± 0.0003 |
370 + | 18 |
133Ba + | EC | 3848.7 ± 1.2 | 80.9979 ± 0.0011 276.3989 ± 0.0012 302.8508 ± 0.0005 356.0129 ± 0.0007 383.8485 ± 0.0012 |
0.329 ± 0.003 0.0716 ± 0.0005 0.1834 ± 0.0013 0.6205 ± 0.0019 0.0894 ± 0.0006 |
37 + | |
152Eu + | EC | 4941 ± 7 | 121.7817 ± 0.0003 344.2785 ± 0.0012 778.9045 ± 0.0024 964.072 ± 0.018 1085.837 ± 0.010 1112.076 ± 0.003 1408.013 ± 0.003 |
0.2841 ± 0.0013 0.2658 ± 0.0012 0.1296 ± 0.0006 0.1462 ± 0.0006 0.1013 ± 0.0006 0.1340 ± 0.0006 0.2085 ± 0.0009 |
37 + | |
241Am | α | 157850 ± 230 | 26.3446 ± 0.0002 59.5409 ± 0.0001 |
0.0240 ± 0.0003 0.3578 ± 0.0009 |
37 | |
* Other activities are available on request. The nominal value of the radionuclide activity in the source within the specified limits is determined by the consumer when the source is ordered. Tolerance on activity ±20 % ** Exposure dose rate - equivalent dose rate of photon radiation at a distance of 0.1 m. Recommended working life and intercalibration interval of the multinuclide source are determined by the shortest working life and interval of individual radionuclides in the source. When supplying sources with a higher activity of the radionuclide, in agreement with the customer, a reduction of recommended working life is allowed. Limit of error of measuring result for confidence probability р = 0,95: 7, 3 or 1.5 % — at certification, attestation in accredited metrological centers or attestation as working standards of the 0-th category, respectively. ISO classification: С35242 |
Sources OSGI-P are used for activity and energy calibration of gamma-spectrometric equipment, testing of ionization chambers and scintillation counters. Rugged design suitable for calibration and field testing.
Source is a transparent disc with diameter 25 mm and height 3 mm. Radionuclide is deposited on active part made of light ceramic pellet. Active part is mounted in center of acrylic capsule and sealed with epoxy resin. Dimensions of active part are 2 mm in diameter and 1 mm in height.
Radionuclide | Half-life, days | Photon energy, keV | Photons per 100 disintegration | Activity, kBq * | Impurities, % ** | Coefficients of the photon radiation transmission *** | ||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
22Na | 950.57 ± 0.23 | 1274.54 | 0.99940 ± 0.00014 | 5 – 1000 | ≤ 0.01 | 0.988 | ||||||||
54Mn | 312.29 ± 0.26 | 834.848 | 0.999746 ± 0.000011 | 5 – 1000 | ≤ 0.15 | 0.985 | ||||||||
57Co | 271.80 ± 0.05 | 122.061 | 0.8551 ± 0.0006 | 5 – 3700 | ≤ 0.15 | 0.967 | ||||||||
60Co | 1925.23 ± 0.27 | 1332.5 | 0.999826 ± 0.000006 | 5 – 3700 | ≤ 0.1 | 0.988 | ||||||||
137Cs | 10990 ± 40 | 661.66 | 0.8499 ± 0.0020 | 5 – 1000 | ≤ 0.05 | 0.983 | ||||||||
241Am | 157850 ± 230 | 59.537 | 0.3578 ± 0.0009 | 5 – 370 | ≤ 0.05 | 0.962 | ||||||||
* The nominal value of the radionuclide activity in the source within the specified limits is determined by the consumer when the source is ordered. Tolerance on activity -10 ÷ +30 %. ** The ratio of the total activity of impurity radionuclides to the activity of the basic radionuclide in the source. *** Coefficients of the photon radiation transmission in OSGI-P source:
ISO classification: С34343 Recommended working life
|