OSGI-P sources are used for efficiency and energy calibration of gamma-spectrometric equipment, testing of ionizing chambers and scintillation counters. Robust design is suitable for “in field” calibrations and tests.
Source is a transparent disc with diameter 25 mm and height 3 mm. Radionuclide is deposited on active part made of light ceramic pellet. Active part is mounted in center of acrylic capsule and sealed with epoxy resin. Dimensions of active part are 2 mm in diameter and 1 mm in height.
Radionuclide | Half-life, days | Photon energy, keV | Probabilities (gamma per decay) | Nominal activity, kBq * | Radionuclide impurities, % ** | Coefficients of the photon radiation transmission *** | ||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
22Na | 950.57 ± 0.23 | 1274.54 | 0.99940 ± 0.00014 | 5–1000 | ≤ 0.01 | 0.988 | ||||||||||
44Ti | 21915 ± 400 | 67.884 | 0.930 ± 0.020 | 5–100 | ≤ 0.001 | 0.963 | ||||||||||
54Mn | 312.29 ± 0.26 | 834.848 | 0.999746 ± 0.000011 | 5–1000 | ≤ 0.15 | 0.985 | ||||||||||
57Co | 271.80 ± 0.05 | 122.061 | 0.8551 ± 0.0006 | 5–3700 | ≤ 0.15 | 0.967 | ||||||||||
60Co | 1925.23 ± 0.27 | 1332.5 | 0.999826 ± 0.000006 | 5–3700 | ≤ 0.1 | 0.988 | ||||||||||
65Zn | 243.86 ± 0.20 | 1115.55 | 0.5060 ± 0.0022 | 5–1000 | ≤ 0.2 | 0.987 | ||||||||||
88Y | 106.625 ± 0.024 | 1836.066 | 0.9938 ± 0.0003 | 5–1000 | ≤ 0.1 | 0.991 | ||||||||||
109Cd | 461.4 ± 1.2 | 88.034 | 0.03626 ± 0.00020 | 5–1000 | ≤ 0.001 | 0.965 | ||||||||||
113Sn | 115.09 ± 0.04 | 391.71 | 0.6494 ± 0.0017 | 5–100 | ≤ 0.01 | 0.978 | ||||||||||
133Ba | 3848.7 ± 1.2 | 356.014 | 0.6205 ± 0.0019 | 5–3700 | ≤ 0.001 | 0.977 | ||||||||||
134Cs | 753.5 ± 1.0 | 604.7 | 0.97650 ± 0.00018 | 5–300 | ≤ 0.01 | 0.982 | ||||||||||
137Cs | 10990 ± 40 | 661.66 | 0.8499 ± 0.0020 | 5–1000 | ≤ 0.05 | 0.983 | ||||||||||
139Ce | 137.642 ± 0.020 | 165.86 | 0.7990 ± 0.0004 | 5–1000 | ≤ 0.1 | 0.969 | ||||||||||
152Eu | 4941 ± 7 | 1408.006 | 0.2085 ± 0.0009 | 5–1000 | ≤ 0.15 | 0.988 | ||||||||||
153Gd | 240.4 ± 1 | 97.43 | 0.29 | 5–500 | ≤ 0.2 | 0.965 | ||||||||||
207Bi | 11800 ± 300 | 569.7 | 0.9776 ± 0.0003 | 5–100 | ≤ 0.2 | 0.982 | ||||||||||
210Pb | 81083 ± 80 | 46.54 | 0.0425 ± 0.0004 | 5–200 | — | 0.960 | ||||||||||
228Th | 698.6 ± 0.23 | 2614 | 0.99754 ± 4 | 5–50 | ≤ 0.2 | 0.998 | ||||||||||
241Am | 157850 ± 230 | 59.537 | 0.3578 ± 0.0009 | 5–370 | ≤ 0.05 | 0.962 | ||||||||||
243Am | 2692000 ± 8000 | 74.67 | 0.672 | 5–50 | < 5 | 0.964 | ||||||||||
* The nominal value of the radionuclide activity in the source within the specified limits is determined by the consumer when the source is ordered. Tolerance on activity -10 ÷ +30 %. ** The ratio of the total activity of impurity radionuclides to the activity of the basic radionuclide in the source. *** Coefficients of the photon radiation transmission in OSGI-P source:
Limit of error of measuring result for confidence probability р = 0,95: 7 или 3 % — at certification or attestation in accredited metrological, respectively. ISO classification: С34343 Recommended working life
|