Mössbauer sources: Cobalt-57 / Co-57

57Co
Radionuclide characteristics
Half-life, days: 271.80 ± 0.05
Photon energy:

Fe XKα26.39084 keV (33.2±0.5%)
Fe XKα16.40384 keV (16.8±0.3%)
Fe XKβ7.058-7.108 keV (7.1±0.2%)
γ114.41295±0.00031 keV (9.15±0.17%)
γ2122.06065±0.00012 keV (85.51±0.06%)
γ3136.47356±0.00029 keV (10.71±0.15%)
Serial sources: 57Co active part is prepared by electrodeposition of high-purity carrier-free 57Co onto a thin (thickness: 6 µm or less, diameter: 4-8 mm) metal matrix followed with controlled annealing process. Standard matrix is rhodium. Active part is placed into titanium alloy holder with brazed beryllium window and sealed with laser welding. In case of ordering a source for liquid Helium temperature application additional notification is required.
Point sources: The source is a beryllium capsule with active part inside, inserted into a titanium alloy holder. The holder is a single-piece element with truncated cone shaped front end and threaded tail piece. Source active part is a metal matrix with thermally diffused 57Co isotope. Source tightness is achieved by glued and welded joints. Point sources are not suitable for use at liquid Helium temperature.
Serial sourcesPoint sources

Please, pay attention, types of the capsules have been changed.

Code Nominal
activity *
Capsule type DxH, mm d, mm
mCi MBq
MCo7.121 5 185 2 14x14 8
MCo7.111 5 185 1 11.2x13 8
MCo7.161 5 185 6 6x13 4
MCo7.122 10 370 2 14x14 8
MCo7.112 10 370 1 11.2x13 8
MCo7.162 10 370 6 6x13 4
MCo7.123 25 925 2 14x14 8
MCo7.113 25 925 1 11.2x13 8
MCo7.124 50 1850 2 14x14 8
MCo7.114 50 1850 1 11.2x13 8
MCo7.125 100 3700 2 14x14 8
MCo7.115 100 3700 1 11.2x13 8

* Tolerance: ±10%

ISO classification: C54243

Recommended working life: 10 years

Line width: 0.110-0.135 mm/s

Other activities are available on request

In case of ordering a source for liquid Helium temperature application additional notification is required

 

Main / Products / Mössbauer sources
   News
06.19.2018
Mössbauer Spectroscopy in Materials Science Conference, 2018 will take place in Prague (Czech Republic) from June 25 to June 28...

06.05.2018
During The International Conference on the Applications of the Mössbauer Effect (ICAME-2017) which was held in September, 2017 in St. Petersburg, Russia, 10 pcs. of Mössbauer sources shielding grippers were drawn by RITVERC...

05.31.2018
The anniversary celebrations devoted twenty-fifth birthday of our company passed May, 18th of this year. Shareholders, employees of Ritverc and the invited visitors, have gone to short travel across Neva river by “City-blues” steam-ship...

05.22.2018
Ritverc will take part in the meeting of the ICRM-LSC work group. The purpose of this meeting is to discuss issues of liquid scintillation spectrometry...

© 2003-2018 RITVERC GmbH
Go
phones: (+7 812) 297 44 63
(+7 812) 297 22 69
Call me back
fax: (+7 812) 297 58 87
e-mail: info@ritverc.com